The StatStrip Xpress®2 professional meter is intended for use in applications that do not require data interfacing and connectivity. Xpress2 provides the same performance and features as the StatStrip hospital connectivity meter, including:
- Only glucose meter technology FDA cleared for use with critically ill patients
- No known clinical interferences–over 8,000 medications tested in 1,698 critically ill patients
- Test strip ejector for improved operator safety
Weight: ....................................... 75 g (2.77 oz )
Size: ......................... 98 mm x 61 mm x 23 mm
(3.9 in x 2.4 in x 0.9 in)
Data Storage:
Patient & QC Tests:.........400 tests total (FIFO)
Operating Ranges:
Temperature:...............5˚C- 40˚C (41˚F-104˚F)
Altitude:...................Up to 4,572 m (15,000 ft)
Humidity:..........10% to 90% relative humidity
Haematocrit: ......................................20%-65%
Battery Information:
Type: ........................................ 2 AAA batteries
Features: ........................................ Replaceable
Life: ................................... Minimum 600 tests
Additional Features:
• LCD colour display
• Large numeric display (30 mm)
• Traditional QC with target values assigned to
QC materials
• Units of measure based on meter ordered
(mg/dL or mmol/L models)
• Automatic shut-off when not in use
• Automatic sample detection and analysis
• Automatic sample counting with date/time
stamp for data tracking
Optional Carrying Case:
Size: ................... 356 mm x 279 mm x 102 mm
(14 in x 11 in x 4 in)
Weight:..........................................75 g (0.17 lb)
StatStrip® Glucose Strips
Test Measured:...........................Blood glucose,
haematocrit-corrected
Test Reported:.......................................Glucose
Test Time: ..........................................6 seconds
Test Strip Volume: .................................. 1.2 µL
Test Methodology: ................Electrochemistry
Sample Types & Operating Modes:
Whole Blood: ........................ Arterial, Venous,
Fingerstick Capillary, Neonatal
Heel Stick, Neonatal Arterial
Glucose Measurement Range:
0.6-33.3 mmol/L (10-600 mg/dL)
Interferences Eliminated:
Haematocrit, Ascorbic Acid, Uric Acid,
Paracetamol, Bilirubin, Maltose,
Galactose, Oxygen, N-acetylcysteine
Test Strip Stability:
30 months from date of manufacture
6 months open-vial stability
StatStrip® Ketone Strips
Test Measured:..............................Blood ketone,
haematocrit-corrected
Test Reported: ..........................................Ketone
Test Time: ..........................................10 Seconds
Test Strip Volume: .....................................0.8 µL
Test Methodology: ...................Electrochemistry
Sample Types & Operating Modes:
Whole Blood: ......................... Venous, Capillary
Ketone Measurement Range:
0.1-7.0 mmol/L
Interferences Eliminated:
Haematocrit, Ascorbic Acid, Uric Acid,
Paracetamol, Bilirubin, Oxygen
Test Strip & QC Stability:
24 months from date of manufacture
3 months open-vial stability
Certifications & Compliance:
Nova Biomedical complies with FDA Quality System Regulations and is certified to meet
EN ISO 13485:2016.
Tested According to: EN 61010-1:2010, EN 61010-2-101:2017, EN 60825-1:2014.
FDA Labeling:
For in-vitro diagnostic use